1.4.6 Environmental conditions

The device is intended for use in rooms with a controlled environment.
The device must not be exposed to adverse/harsh environmental conditions.

1.5. Warnings/caution

Warning:
The general warning symbol indicates a potentially dangerous situation that can lead to
serious injuries.
Do not use in a magnetic resonance environment.
There is a risk of ignition of gases if the device is operated in the presence of flammable
mixtures or mixtures of medicinal products and air or oxygen or nitrous oxide!
The device must not be operated in rooms in which flammable mixtures or mixtures of
pharmaceuticals and air or oxygen or nitrous oxide are present, e.g. operating rooms.
Electric shock!
The housing of the examination lights may only be opened by authorized persons.
Damage to the device due to a fall or strong ESD influence!
If the device is not functioning, it must be returned to the manufacturer for repair.
The device must be used in a controlled environment.
The device must not be exposed to harsh environmental conditions.
Caution!
The caution symbol indicates a potentially dangerous situation that can lead to minor or
moderate injuries. It can also be used to warn of unsafe practices.
The faultless and safe functioning of these examination lamps can only be guaranteed if
Riester original parts and accessories are used.
Old electronic devices must be disposed of in accordance with the institutional guidelines
for the disposal of expired devices.
Cleaning frequency and sequence must comply with the cleaning regulations of non-sterile
products in the respective facility. Cleaning/disinfection instructions in the instructions for
use must be observed.
We recommend unplugging the power supply before cleaning or disinfecting.
Clean and disinfect the examination lamps carefully so that no liquid penetrates the in-
terior.
Never place the examination lamps in liquids!
The examination lamps are delivered in a non-sterile condition. Do not use ethylene, oxide
gas, heat, autoclaves or any other harsh methods to sterilize the device. The devices have
not been approved for mechanical reprocessing or sterilization. This leads to irreparable
damage!
The patient is not the intended operator.
The product may only be used by qualified personnel.
Qualified personnel are doctors or nurses in hospitals, medical facilities, clinics and me-
dical practices.

1.6. User Responsibility

Caution!
User Responsibility
It is your responsibility to:
The user must check the integrity and completeness of the examination lamps before each
use. All components must be compatible with each other.
19