Priming Function
In order to facilitate the priming the pump is equipped with PRIMING function. It's recommended to open the bleed air
by-pass on pump head (for pumps with it) during priming. The air bleed nipple must always be connected with the return
hose to the suction tank.
The activation of this function can occur at any operating menu in the following way:
Press and hold START/STOP button
After about 3 seconds the pump starts delivering injections at the priming frequency of 150 imp/min, as long as
the operator does not release the START / STOP button. When you press the button the following screen
appears on the display:
Once you release the button, the pump stops 10 sec. countdown starts. At the end the pump switches back to
the previous operating mode, in stand-by condition. Pressing the START / STOP button the pump goes in
running mode.
During the countdown pressing and holding down START/STOP the pump starts to give injections to reach the
right dosing condiction to existing system pressure. We recommend that you use this procedure to return the
pump in optimal working condition especially if it operates at any proportional mode. During this phase, you
should close of the bleed valve on the pump body. Once you release the button, the pump will return to stand-
by mode.
Fig.5 – Bleed operation for priming