OUT_WD2@m
Watchdog mode 2: When a WD2 event occurs, the speed set point will be set to the WD safety
set point speed. The PC 2 warning is displayed. The WD2 event can be reset with OUT_WD2@0-
resetting also blocks the watchdog function. Set the watchdog time to m (20...1500) seconds, with
echo of the watchdog time. This command starts the watchdog function and must be sent within
the set watchdog time.
RESET
Switching off the device function.
START_4
Starting the device (remote) function
STATUS_4
Display of status:
10: Manual operation without fault
11: Automatic operation Start (without fault)
12: Automatic operation Start (without fault)
Note: when disconnecting the RS 232 or USB cable while the device is running, the device stays in PC mode. To reset to
function without PC, switch the device off and on again.
PC 1.1 Cable:
This cable is required to connect RS 232 interface (G) to a PC.
1
6
2
7
3
8
4
9
5
1
RxD 2
TxD 3
4
GND 5
6
RTS 7
CTS 8
Fig. 4
9
USB 2.0 cable A - B:
This cable is required to connect USB interface (H) to a PC.
4
1
3
2
USB B
Fig. 5
< 0: error code:
- 1: error 1
- ... (see „Error codes" table)
5
9
4
8
3
7
2
6
1
1
2 RxD
3 TxD
PC
4
5 GND
6
7 RTS
8 CTS
9
1
2
3
4
USB A
16