Always make sure that the flow control knob is in the correct position and
that it is not in a position between two values. In such case the flow control
head will not deliver the correct flow of the medical gas.
If the flow control knob stops at the maximum flow rate or at the "0" value,
do not try to exert excessive force when turning it.
EN
The oxygen flow rate must be prescribed and provided by a clinically trained
user.
AFTER THE END OF THERAPY
Turn off the cylinder valve by turning the hand wheel in a clockwise direc-
1.
tion to the "stop" position.
Vent gas pressure out of the connected devices.
2.
Set the flow control knob to the "0" (OFF) value when there is no venting
3.
anymore – ensure that the flow control knob engages correctly.
Disconnect the hose/humidifier from the flow outlet.
4.
USE OF THE PRODUCT PRESSURE OUTLET
Ensure that the flow control knob is set to the "0" value (only for products
1.
with flow metering devices).
Ensure that the accessory IS NOT connected to the pressure outlet.
2.
Slowly open the cylinder valve by turning the hand wheel in an anticlock-
3.
wise direction with approx. 1 to 1½ turns.
Sudden opening could result in a danger of fire or explosion arising from
oxygen pressure shocks. Insufficient opening of the shut-off valve could re-
duce actual flow delivered.
Connect the accessories to the pressure outlet.
4.
AFTER THE END OF THERAPY
Turn off the cylinder valve by turning the hand wheel in a clockwise direc-
1.
tion to the "stop" position. Do not use excessive force.
Vent gas pressure out of the connected devices.
2.
Disconnect the quick-coupler adapter from the pressure outlet.
3.
6.3. AFTER USE
Turn off the cylinder valve by turning the hand wheel in a clockwise direc-
1.
tion to the "stop" position. Do not use excessive force.
Ensure that the flow control knob is set to the "0" (OFF) value until the flow
2.
control device reaches the correct position (only for products with flow
metering devices).
Make sure that the pressure indicator/sensor does not indicate any residual
3.
pressure.
Disconnect all devices connected from the user outlets.
4.
10/115