Hach PAH500 Manuel De L'utilisateur page 3

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 79
Table of Contents
1 Specifications
on page 3
2 General information
on page 4
3 Installation
on page 10
Section 1 Specifications
Specifications are subject to change without notice.
Specification
General
Dimensions
Cable length
Materials
Weight
Mounting
(optional kit available)
Plumbing
(user supplied)
Protection class
Pollution degree
Electrical installation category
Power requirements
Operating ambient temperature 5 to 55 °C (41 to 131 °F), 100% relative humidity maximum, condensing
Storage temperature
Certification
Warranty
Software requirements
SC1000 Controller
Sample requirements
Pressure
Flow
Temperature
4 Operation
5 Maintenance
6 Troubleshooting
Details
Sensor (Ø x L): 76.2 x 145.1 mm (3.0 x 5.7 in.)
Digital gateway (Ø x L): 34.6 x 182.4 mm (1.36 x 7.18 in.)
Cable between the sensor and digital gateway: 1 m (3.3 ft); flame rating IEC
60332-1-2; UL94 VW-1
Sensor enclosure: Titanium and NBR (nitrile butadiene rubber)
Sensor wetted materials: Quartz glass, fluorosilicone and titanium
Digital gateway: Ryton (PPS); flame rating UL94 V0
Sensor: 1000 g (35 oz)
Digital gateway: 145 g (5 oz)
Wall mount
Tubing: 6 mm (¼-inch) ID recommended
Fittings: ¼-18 NPT 6 mm (¼-inch) ID fittings
III
3
I
Sensor: 12 VDC supplied by digital gateway
Digital gateway: 12 VDC supplied by SC1000 Controller, 1.8 W
-20 to 60 °C (-4 to 140 °F), 0 to 95% relative humidity, non-condensing
CE approved
DNVGL-CG-0339
MEPC.259(68)
FCC SDoC - FCC Part 15B, Class A, when used with SC1000 Controller
1 year (EU: 2 years)
V3.37 or later
7 bar (101 psi) maximum
Continuous sample flow is necessary for accurate measurements.
Recommended 1 L/min of sample flow.
0 to 50 °C (32 to 122 °F)
on page 14
on page 22
on page 23
English 3

Publicité

Table des Matières
loading

Table des Matières