Function Heating

The temperature of the heating plate of the instrument is kept
constant by a control circuit. In addition, it is monitored by a safety
circuit. The required temperature sensor (a thermo-couple) is
installed in the heating plate. The RH basic unit has a stainless
steel surface plate with a heating output of 400 Watt.
The temperature of the heating plate can be adjusted on the
„Temp" turn dial (E) with a corresponding scale from 0 to 6 (F).
The scale values from 0 to 6 correpond approximately to area tem-
peratures up to a max. of 350 °C. When the setting is at 0, the
device does not heat the medium up at normal room temperatu-
re.
Mounting - Support rod H 16 V
The H 16 V support rod can be screwed into the threading on the
right behind the housing as follows:
• Remove the plastic cap and washer from the support rod and
screw off the hex nut.
• Place the washer and hex nut on the beginning of the threa
ding on the housing and hold them in place.
• Screw the support rod into the nut and threading of the hou
sing and counter-tighten the hex nut.

Maintenance and Cleaning

The IKA RH basic 1 is maintenance-free. It is subject only to the
natural wear and tear of components and their statistical failure rate.
When ordering spare parts, please give the manufacturing
number shown on the type plate, the machine type and the
name of the spare part.
RH b2dKT
All manuals and user guides at all-guides.com
Please send in equipment for repair only after it has been cleaned
and is free from any materials which may constitute a health
hazard. Please tell us if you have used any materials which are
dangerous or which may constitute a health hazard. Use only cle-
ansing agents which have been approved by IKA to clean IKA
devices.
To remove use:
Dyes
isopropyl alcohol
Construction materials water containing tenside / isopropyl alcohol
Cosmetics
water containing tenside / isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information from
IKA. Wear the proper protective gloves during cleaning of the
devices. Electrical devices may not be placed in the cleansing
agent for the purpose of cleaning.
Before using another than the recommended method for cleaning
or decontamination, the user must ascertain with the manufactu-
rer that this method does not destroy the instrument.

Accessories

RS 1
set of stirring bars
H16V support rod
H 15
bath top
RSE
H 36
holding rod
H 44
cross sleeve
Recommended IKA-stirrers
TRIKA-stirrer PTFE, triangular, length 40 mm
IKAFLON-stirrers PTFE
Diameter 8 mm, lengths 25, 30, 40 and 50 mm