Remote Ctrl
Remote control operation. The power unit can be remote-controlled via an external voltage or
external poti. The remote settings can be performed for voltage and current. The front controls
are inactive.
Set
Settings operation. The three preset slots can be programmed freely. Select the memory slot with
the "RECALL" (15) slider and make the settings using the controls (7, 8).
The separate operating modes are described in more detail in the following.
d) Added functions
The power supply is auto-zeroed every time you turn it on. In case you need to zero the unit during operation and do
not want to restart it, zero it manually.
1. Press and hold the VOLTAGE control knob for approx. 30 s to enter the MENU mode. "CCO" and "no" are dis-
played.
2. Rotate the CURRENT control knob until "CCO" and "YES" are displayed.
3. Press the CURRENT control knob once to zero the unit. "YES" lights up in the display to confirm successful zero-
ing.
4. Press the VOLTAGE control knob to exit the MENU mode.
e) Normal operation
In normal operation, the power unit can be operated through the front controls.
1. Ensure that the "MODE" slider is in the "Normal" position.
2. Remove any connected consumers from output (9 and 10 or 11 and 12).
3. Switch on the power unit at the power switch (1). The display (2 and 3) lights up, and after a
short selftest, the current and voltage displays appear.
Set the current limitation before setting any voltages. If the current value is too high,
your connection lines can be damaged; if it is too low (<1 A), the output voltage can
be limited.
35