E
Serial number:
The unit bears a serial number
identification.
Power consumption:
Maximum 4 W
Power supply:
Battery type: 4 x 1.5 V AA alkali manganese (LR 6) or
lithium (FR 6) round cells
Optional: Stabilised type A1 mains adapter, output 6
VDC, minimum 0,6 mA
Cuff:
Cuff UW (for arm circumference 23-43 cm)
Operating conditions:
Ambient temperature 10 to 40 °C, rel. air humidity 15 to
85 % not condensing, air pressure 700 to 1060 hPa
Storage and transport
Ambient temperature -5 to 50 °C, rel. air humidity up to
conditions:
85 % not condensing,
Inflation and defla-
Automatically controlled
tion:
Automatic switch-off
approx. 3 minutes after end of measurement
IP code:
IP 20: Protected against solid foreign particles with a
diameter of more than 12.5 mm, no protection against
water.
Expected service life:
5 years
Classification:
Internal power supply by battery
4. Original spare parts and accessories
The following original spare parts and accessories are available from specialist
dealers:
Cuff 23-43 cm type UW
Part. no. 2401601, PZN 10541010
Subject to technical modifications.
EN-52
Technical information
which provides clear
Mains adapter type A1
Part. no. 2401020, PZN 03558547
Technical information
5. Technical inspection / calibration check
Basically, a metrological inspection is recommended at intervals of 2 years. Howev-
er, professional users in Germany are obligated to comply with the aforementioned
according to "Regulation for Operators of Medical Devices".
This can be implemented either by UEBE Medical GmbH, an authority responsible
for metrology, or authorised maintenance services. For this, please observe your
national provisions.
Upon request, responsible authorities or authorised maintenance services receive a
"Test instruction for metrological inspection" from the manufacturer.
Important: No modifications, e.g. opening the unit (except to replace the batteries)
may be made to this unit without the manufacturer's prior permission.
6. Explanation of symbols
This product complies with the Council Directive 93/42/EC from 5
September 2007 regarding medical devices and bears the mark CE
0123
0123 (TÜV SÜD Product Service GmbH).
Degree of protection against electric shock: TYPE BF
Please observe instructions for use
Storage and transportation conditions
Ambient temperature -5 to +50 °C
85%
Relative humidity up to 85 %, non-condensing
0%
Keep dry
IP20
Protected against solid foreign particles with a diameter of more than
12.5 mm, no protection against water.
E
EN-53