STATUS_X
X=1;4;5
X=16
(old function)
STOP_X
X=1;2;4;5;7
"Watchdog" functions; monitoring of the serial data flow
If, once this function has been activated (see NAMUR commands), there is no retransmission of the command from the computer within
the set time ("watchdog time"), the heating and stirring functions are switched off in accordance with the set "watchdog" function or
are changed to the set target values.
The data transmission may be interrupted by, for example, a crash in the operating system, a power failure in the PC or an issue with the
connection cable between the computer and the device.
"Watchdog" - mode 1
If there is an interruption in data communications (longer than the set watchdog time), the heating and stirring functions are switched
off and ER 2 is displayed.
"Watchdog" - mode 2
If there is an interruption in data communications (longer than the set watchdog time), the speed target value is changed to the WD
safety speed limit and the temperature target value is changed to the WD safety temperature limit value. The warning WD is displayed.
PC 2.1 cable
The PC 2.1 cable is used to connect the 9-pin plug to a computer.
1
6
2
7
3
8
4
9
5
1
RxD 2
TxD 3
4
GND 5
6
RTS 7
CTS 8
9
Display of status
1*: Operation mode A
2*: Operation mode B
3*: Operation mode C
*0: manual operation without fault
*1: Automatic operation Start (without fault)
*2: Automatic operation Start (without fault)
<0: Error code: (-1)
-
1: Error 1
-
... (see table)
-
31:Error 31
-83: wrong parity
-84: unknown instruction
-85: wrong instruction sequence
-86: invalid rated value
-87: not sufficient storage space
Switching off of device - (remote) function The variables set with OUT_SP_X
remain saved. Includes command RMP_STOP.
5
9
4
8
3
7
2
6
1
1
2 RxD
3 TxD
4
PC
5 GND
6
7 RTS
8 CTS
9
Remote
23