8 Adjustment – Easy Calibration
8 Adjustment – Easy Calibration
The instrument is permanently calibrated for aqueous solutions. If it is
determined that the pipette is operating inaccurately or to adjust the in-
strument to work with solutions of varying density and viscosity or with
specially-shaped pipette tips, it can be calibrated using the Easy Cali-
bration Technique.
b.
d.
Operating manual
50
a. Perform a volume check
b. Remove label window (1)
1
c. Using a paper clip or an un-
2
3
d. Slide the red adjustment
and determine the actual
value; see p. 46.
and labeling film (2): Gently
move the clamp and lift it
off.
used pipette tip, remove the
protective film (3) (the pro-
tective film can be dis-
carded).
slider back completely, lift
the volume-setting wheel
(decoupling) and release the
adjustment slider.
Transferpette® S 9974 88 V1