Heating Bath; Commissioning - IKA RV 3 eco Mode D'emploi

Masquer les pouces Voir aussi pour RV 3 eco:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 31
• Mount the washer
Drainage washer
NOTICE
Pay attention to the correct
position of the washer!
Hose system

Heating bath

Refer to the chapter entitled
CAUTION
"Commissioning" in the operating
instructions of heating bath.
Switch on the device
The unit is ready for service when
the mains plug has been plugged
in.
• Description of special condensers
(See products/accessories at www.ika.com)
- RV 10.3 Vertical-intensive condenser with manifold.
This vertical-intensive condenser features a double
jacket design for particularly efficient condensation.
Also available with coating (RV 10.30)
- RV 10.4 Dry ice condenser
Dry ice condenser is for distilling low-boiling solvents.
Cooling by dry ice, no cooling water is required.
Maximum condensation thanks to low temperatures.
Also available with coating (RV 10.40)
- RV 10.5 Vertical-condenser with manifold and cut-off
valve for reflux distillation.
Also available with coating (RV 10.50)
- RV 10.6 Vertical-intensive condenser with manifold and
cut-off valve for reflux distillation
This vertical-intensive action condenser features a
double jacket design for particularly efficient reflux
distillation.
Also available with coating (RV 10.60)
- Connect the water hoses (H
according to the counter-flow principle.
- Install the vacuum connections to the condenser,
condensate flask, vacuum controller with valve and
vacuum pump.
- Always connect the vacuum hose (vac) to the condenser
at the highest point to minimize solvent losses during
suctioning.
- Use standard laboratory vacuum hoses with an inside
diameter of 8 mm and a wall thickness of 5 mm.

Commissioning

Use the power switch to switch on or off the device.
Adjust the speed by turning the speed adjusting knob.
O) to the condenser
2
23

Publicité

Table des Matières
loading

Table des Matières