Error sound
Cause
Continuously
- The speed has not been adjusted to
beeping
0 rpm when starting the device
- PCBA failure
Alternatively
- Overload
beep 1 sec and
- Motor blocked
stop 1 sec
- Defective cable
- PCBA failure
Alternatively
- Temperature inside the device is too
beep 0.5 sec
high
and stop 1.5 sec
- PCBA failure
Alternatively
- Overvoltage
beep1.5 sec and
- Undervoltage
stop 0.5 sec
- PCBA failure
- Desktop power supply failure
Operating voltage range
Rated voltage
Frequency
Rated power without heating bath
Rated power (max.) of switching adaptor
Speed
Speed tolerance
Speed display
Smooth start
Lift
Stroke
Head angle adjustable
Cooling surface
Cooling water flow rate min.
Cooling water flow rate max.
Cooling water pressure max.
Perm. On-time
Perm. ambient temperature
Perm. relative humidity
Protection acc. to DIN EN 60529
Protection class
Contamination level
Weight (no glassware; no heating bath)
Dimensions (W x D x H)
Operation at a terrestrial altitude
Subject to technical changes!

Technical data

Set speed: < 100 rpm
Set speed: ≥ 100 rpm
Effect
Correction
Motor can't run
- Turn the speed adjusting knob and
set the speed to 0 rpm, after which
you may set the required speed val-
ue.
Motor stops running - Reduce the volume in the evaporator
- Contact the service department
Motor stops running - Switch off the device
- Let the device cool down
- Switch on the device
- Contact the service department
Motor stops running - Use the original power supply to
supply the device
- Contact the service department
Vac
Vac
Hz
W
W
rpm
rpm
%
mm
cm
2
l/h
l/h
bar
%
°C
%
kg
mm
m
(100-240) ± 10%
100...240
50/60
40
90
20...300
± 1
± 1
Plastic part + scale
yes
manual
150
0°...60°
1500
30
100
1
100
5...40
80
IP 20
I
2
8
440 x 280 x 590
max. 2000
29