MIETHKE proSA Tools Mode D'emploi page 29

Utilisés pour régler et contrôler les niveaux de pression paramétrés de la valve prosa
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
proSA
®
Tools
Fig. 7: Placing the Verification Tool
Tool
Fig. 8: Pressure level measurement
a) correct
b) wrong
2b. Testing with the proSA
Verification Compass
proSA Verification Compass
Ventricular Catheter
NOTE
Small air bubbles in the Verification Compass
chamber do not impair its function.
Verification Tool
Verification
3. Adjustment process
WARNING
When adjusting the adjustable Gravitational
Unit, care must be taken that the open-
ing pressure is changed by a maximum of
16 cmH
O per adjustment process.
2
Example: The opening pressure for the
adjustable Gravitational Unit on the proSA is
to be changed from 6 to 36 cmH
is only one adjustment process, the rotor will
rotate anti-clockwise (short route) and stop
at the setting 0 cmH
The correct method is an adjustment in two
stages: Initial adjustment from 6 cmH
22 cmH
O and then from 22 cmH
2
36 cmH
O. The rotor rotates clockwise (Fig.
2
9b).
Fig. 9: Rotor rotation during adjustment
a) wrong direction
b) correct direction
NOTE
During preoperative adjustment through
the sterile packaging, the Adjustment Tool
is only placed on the valve. Do not press
the push-button. Excessive pressure with
the Adjustment Tool can damage the valve
housing and impair its function.
If the implantation site is unfavourably cho-
sen or the skin above the valve is too thick,
the adjustable Gravitational Unit may no
longer be able to be adjusted. The valve
then works with unchangeable pressure
levels. The patient should be made aware
of this risk.
Swelling of the skin may make adjustment
difficult for a few days after the operation.
EN
O. If there
2
O (Fig. 9a).
2
O to
2
O to
2

Publicité

Chapitres

Table des Matières
loading

Table des Matières