3.
Wire size: 18 AWG (0.832 mm
24 AWG (0.205 mm
OUTPUT
The four dry-contact normally open relay outputs can switch up to 250 VAC or 30 VDC voltages at 1A
maximum.
The relay outputs are not polarity-sensitive since they are dry contact outputs.
An example of wiring to the outputs is given below:
COMPLIANCE
Mark
Standard
This product complies with the applicable harmonized standards of EU
Directives 2011/65/EU (RoHS) and 2014/30/EU (EMC). The EU Declaration of
Conformity is available online at www.ohaus.com/ce.
FCC Note
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case
the user will be required to correct the interference at his own expense.
Industry Canada Note
This Class A digital apparatus complies with Canadian ICES-003.
2
) maximum
) minimum
2
Notes:
1.
Dry contact relays.
2.
Relay contact rating:
AC: 24-250 VAC, 1.0 AMP. INTO resistive load.
DC: 5-45 VDC ,1.0 AMP. INTO resistive load.
Maximum switching power: 250VA, 30 W.
3.
Maximum output circuit current = 3 AMPS.
4.
All inductive loads must be suppressed.
5.
Wire size: 18 AWG (0.832 mm
2
) Maximum
24 AWG (0.205 mm
2
) Minimum
EN-13