EN-12
10. Fasten the cap nut of the M25 gland.
CONNECTION
INPUT
The discrete I/O interface enables other devices such as PLCs to provide the trigger voltage (typically
12 VDC or 24 VDC, maximum 30 VDC) to turn the inputs "on".
An example of wiring to the inputs with the +V to the common is shown:
Note:
1.
Voltage: 5~30 VDC, 10 mA maximum current. Voltage polarity cannot be reversed.
2.
Do not bundle input wiring with power wiring or any other high energy cables.
Rubber Seal
Cap nut of M25 gland