Table 3 - Guidance and manufacturer's declaration – electromagnetic immunity – for ME
equipment and ME systems that are not life-supporting
The EVOLUTION 3E is intended for use in the electromagnetic environment specified below. The customer or the user of
the EVOLUTION 3E should assure that it is used in such an electromagnetic environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
(1) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the EVOLUTION 3E is used exceeds the applicable RF
compliance level above, the EVOLUTION 3E should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the EVOLUTION 3E.
(2) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Guidance and manufacturer's declaration – electromagnetic immunity
Immunity test
3 Vrms
3 Vrms
150 kHz to 80 MHz
10 V/m
3 V/m
80 MHz to 2,5 GHz
Immunity test
Portable
equipment should be used no closer to any part
of the EVOLUTION 3E, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.17 √P
d = 1.17 √P 80 MHz to 800 MHz
d = 2.33 √P 800 MHz to 2,5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey (1)
should be less than the compliance level in each
frequency range. (2)
Interference may occur in the vicinity of equipment
marked with the following symbol:
Immunity test
and
mobile
RF
communications
#65073 - B - 01.2017
MORIA