Table 1 - Guidance and manufacturer's declaration – electromagnetic emissions – for all
ME equipment and ME systems
Guidance and manufacturer's declaration – electromagnetic emissions
EVOLUTION 3E is intended for use in the electromagnetic environment specified below. The customer or the user of
EVOLUTION 3E product should assure that it is used in such an electromagnetic environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 2 - Guidance and manufacturer's declaration – electromagnetic immunity – for all
ME equipment and ME systems
The EVOLUTION 3E is intended for use in the electromagnetic environment specified below. The customer or the user of
the EVOLUTION 3E should assure that it is used in such an electromagnetic environment.
Immunity
test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60
Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level
4
MORIA
#65073-B- 01.2017
Compliance
Group 1
Class A
Class A
Complies
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601
test level
±6 kV contact
±8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT (>95 % dip in
UT) for 0,5 cycle
40 % UT (60 % dip in
UT) for 5 cycles
70 % UT (30 % dip in
UT) for 25 cycles
<5 % UT (>95 % dip in
UT) for 5 s
3 A/m
Electromagnetic environment – guidance
The EVOLUTION 3E uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The EVOLUTION 3E is suitable for use in all establishments
other than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Compliance
level
±6 kV contact
±8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
±1 kV mode differential
±2 kV mode common
0 Vac during 10 ms
92 Vac during 100 ms
161 Vac during 1 s
0 Vac during 5 s
3 A/m for 50Hz and for
60Hz
Electromagnetic environment
guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
EVOLUTION 3E requires continued
operation during power mains
interruptions, it is recommended that
the EVOLUTION 3E be powered
from an uninterruptible power supply
or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial
or hospital environment.