QUICK REFERENCE GUIDE: TROUBLEShOOTING
ACTION TAKEN
SympTOm
1.
Check that the Main Power Switch is in the ON (I) position, and that the detachable power cord is fully seated in the
System will not
receptacle on back of system.
operate: No Power
2.
Check that the system's power cord plug is fully seated in an appropriate AC wall outlet.
ON Indicator
3.
Check that the wall outlet is functional.
1.
If the office has more than one Tap-On
System will not
used. With a handpiece and insert installed, depress the Tap-On
operate: Power ON
dispense water. If none of the Tap-On
Indicator is illuminated
2.
Resynchronize one Tap-On
1.
Assure that handpiece lavage control is properly adjusted.
System operates:
2.
Check for clogged insert. Replace insert if necessary.
No water flow
3.
Check that dental office water supply valves are open.
to insert tip or
4.
If the system is connected to DualSelect Dispensing System, check that fluid level in the selected bottle is sufficient.
Make sure valves are open when using external water source.
handpiece
5.
Check that the water line filter is clean. Replace filter, if needed.
overheats
1.
Check that the Power Control is not in Rinse Mode.
System operates:
2.
Check the insert for damage and that it is properly installed in the Handpiece.
No insert
3.
Check that the handpiece is properly installed to the cable assembly.
cavitation
4.
Verify that the soft nozzle grip is flush with the hard plastic of the insert port.
5.
Turn the system's Main Power Switch OFF, (O) position. Wait 5 seconds and turn the system back ON.
6.
If problem still exists, replace both "AA" batteries in Tap-On
connect Auxiliary Tap-On
1.
Fast blinking (3 blinks per second) – Indicates improper set-up
System operates:
A. If insert is in the handpiece, remove. Verify the handpiece is properly seated and hold the Tap-On
Service
indicator
B. Attach a NEW handpiece and depress Tap-On
blinking
C. Install and fully seat an insert into handpiece. Depress Tap-On
D. Install and fully seat a NEW insert in handpiece and depress Tap-On
2.
Slow blinking (1 blink per second) - The system is not operating within factory specifications.
A.
B.
C.
D.
1.
Ensure that the base unit has adequate ventilation and is not near a heat source (i.e. radiator, h eat lamp, sunlight or other
System operates:
heat producing operatory equipment).
Service indicator
2.
Turn Main Power Switch OFF (O) position. Allow system to cool for 10 minutes and turn system to the ON (I) position.
illuminated
Verify light is not illuminated.
3.
If light is still illuminated, refer to Section 11.2 Technical Support and Repairs to have unit serviced as soon as possible.
System operates:
1.
Check that there is no insert in the handpiece.
Purge mode will not
2.
Check that Handpiece is properly installed to the cable assembly.
function - icon flashing
TM
Foot Pedal, test each to ensure that the proper Tap-On
TM
Foot Pedals operate the system, continue to the next step.
Foot Pedal to the system (see Section 7.10 Tap-On
TM
Foot Pedal Cable.
TM
2 seconds. If blinking stops, the system is ready for use. If blinking remains, continue to the next step.
for use. Discard the old handpiece or return if within warranty. If blinking remains, continue to the next step.
system is ready for operation. If blinking remains, continue to the next step.
system is ready for use. Discard old insert or return if within warranty. If blinking remains, refer to Section 11.2
Technical Support and Repairs to have unit serviced as soon as possible.
Remove insert.
Turn Main Power Switch OFF, (O) position. Wait five seconds. Turn unit ON, (I) position.
Operate Purge function.
If service indicator still blinks, refer to Section 11.2 Technical Support and Repairs to have unit serviced as soon
as possible.
23
TM
Foot Pedal to the first position. The system should
Foot Pedal Synchronization).
TM
Foot Pedal with new "AA" batteries (Refer to Section 7.9) or
TM
Foot Pedal for 2 seconds. If blinking stops, the system is ready
TM
Foot Pedal for 2 seconds. If blinking stops, the
TM
Foot Pedal for 2 seconds. If blinking stops,
TM
TM
Foot Pedal is being
TM
Foot Pedal for