8 2 Diagnostic Display Indicators and Control
Service Indicator
I lluminates when the system
is not functioning properly.
T his display has three
distinct modes.
• A fast blink (3 blinks per
second) indicates an
improper set-up.
• A slow blink (1 blink per
second) means the
system is operating out
of factory specifications.
• A steady light indicates
the system is overheating.
R efer to section 11.1 for
Troubleshooting guidelines.
Power Indicator
Illuminates (3 sec. delay)
when the Main Power ON/
OFF Control Switch is ON
("I" position).
Low Battery
Indicator
I lluminates when
the Tap-On
Foot
TM
Pedal battery power
is a pproaching end of
life. Replace batteries
as instructed in
section 7.9.
Purge Control
I lluminates when the Purge function is activated.
T o activate Purge, remove insert from the
handpiece, turn the handpiece lavage control
to maximum water flow and press the Purge
button. Water will purge through system for
2 minutes. To deactivate mode during the 2
minute cycle, press Purge button again or press
Tap-On
Foot Pedal.
TM
The Purge Control is also used during the Tap-
On
Foot Pedal Synchronization process. See
TM
section 7.10 for more details.
12
Boost Indicator
Illuminates when the
Boost Mode has
been activated with
the Tap-On
Foot Pedal.
TM
Turbo Mode
When pressed, extra
power is delivered to the
system, up to 25%. This
will increase the stroke
of the tip. Turbo power
remains on until the
button is pressed again
or unit is turned off.
(Purple arrows illuminate
when Turbo Mode is on).