8 3 handpiece / Cable
8 4 Cavitron 30K Ultrasonic Inserts
The many styles of Cavitron and Cavitron Bellissima 30K Ultrasonic Inserts are easily interchangeable for various procedures
and applications. See enclosed literature for specific information.
O-Ring
Provides seal for handpiece coolant.
O-ring should be replaced when worn.
Insert Tip
Shape and size of tip
determines access and
adaptation. Preheated
Lavage directed to tip.
Insert Marking
Finger Grip
Manufacturer,
Date (YDDD=Single digit
year and triple digit day
of year), Frequency,
Type, Tip Lot Number
(if applicable).
Lavage Control
Turn the Lavage Control to select flow rate during system
operation. Flow rate is based on a scale from 1 to 6. Turn
clockwise toward 6 to increase flow at insert tip. Turn
counter-clockwise toward 1 to decrease flow. The flow rate
through the handpiece also determines the temperature of
the lavage. Lower flow rates produce warmer lavage. Higher
water flow rates produce cooler lavage.
If the handpiece becomes warm, increase the flow rate.
With experience the Dental Healthcare Professional will be
able to determine the best flow rate setting for optimum
operating efficiency and patient comfort.
Swivel Feature
Reduces cable drag as handpiece rotates during procedures.
Steri-Mate
Grip accessory (not shown)
®
The Steri-Mate
®
Grip is designed to provide an ergonomic
and comfortable grasp of the handpiece. The grip is
sterilizable and is available in several different colors as an
accessory for your Steri-Mate Handpiece. See installation
instructions provided with the grip.
Connecting Body
Transfers and amplifies mechanical
motion of stack to insert tip.
Magnetostrictive Stack
Converts energy provided by the handpiece into
mechanical oscillations used to activate the insert
tip.