Prob = 1
rEF0 CHF1
....
rEF1 CH1
010.0000
....
0.0000
If a channel is not calibrated, the display show :
CAL 1 Err
CAL 2 Err
CAL Err
If a probe is not connected, the display show :
no Prob 1
no Prob 2
no Prob
If a point per point correction is activated while calibrating the unit, it is deacti-
vated but not erased. By pressing the Preset key when the channel is in abso-
lute mode (Abs) the display briefly indicates Abs.
Prob = 2
Calibration of channel 1 or
Calibration of channel 2 or
Calibration of 2 channels (1 and 2)
Place the probe on reference 0
Acquisition of the reference 0
Insert the reference value (slip gauge value) using
The key
reference 1, then press Preset key.
Acquisition of the reference 1
The display return to the value of channel 1
An error during calibration displays the message
<rEF1 Err> or <rEF2 Err> for channel 2
according to the active current function
according to the active current function
10
and
then place the probe on
Prob = 1.2