Guidance and manufacturer's declaration – Electromagnetic immunity
The HemoCue systems are intended for use in the electromagnetic environment specified below. The customer or
user of the HemoCue systems should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for power supply
transient/burst
lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential mode
IEC 61000-4-5
Voltage dips, short
<5 %U (>95 % dip in U)
interruptions and
for a 0.5 cycle
voltage variations
on power supply
40 %U (60 % dip in U)
lines
for 5 cycles
IEC 61000-4-11
70 %U (30 % dip in U)
for 25 cycles
<5 %U (>95 % dip in U)
for 5 seconds
For explanation of U see
NOTE 1.
Compliance level
Electromagnetic environment – guidance
±6 kV contact
Floors should be wood, concrete or cera-
mic tile. If floors are covered with synthetic
±8 kV air
material, the relative humidity should be at
least 30 %.
2 kV for power
Mains power quality should be that of a typi-
supply lines
cal commercial or hospital environment.
±1 kV for input/
output lines
±1 kV differential
Mains power quality should be that of a typi-
mode
cal commercial or hospital environment.
<5 %U (>95 % dip
Mains power quality should be that of a typi-
in U) for a 0.5 cycle
cal commercial or hospital environment.
40 %U (60 % dip
If the user of the HemoCue systems requires
in U) for 5 cycles
continued operation during power mains
interruptions, it is recommended that the
70 %U (30 % dip
HemoCue systems be powered from an unin-
in U) for 25 cycles
terruptible power supply or a battery.
<5 %U (>95 % dip
in U) for 5 seconds
56