Vx0100 - Metrix BioTest VX0003 Notice De Fonctionnement

Testeur de champs électriques bf
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 5
Chapter II

3.2. VX0100

The tester measures low frequency electric fields.
Measurements can be made in the following bandwidths:
10 Hz - 3 kHz
3 kHz - 100 kHz
10 Hz - 100 kHz
Warning: moving the tester too quickly will cause
detection of parasite electric fields.
It has the following additional functions:
Sound signal:
Sound
indicating
levels. The sound emitted is proportional to the
measured field intensity. This is switched on
and off using the
Hold:
Keeps the measured value until "HOLD" is
pressed again.
Activate and deactivate by applying a short
push on the "HOLD" key.
Filter:
Repeat changes by pressing the "FILTER"
key.
Measurements in the bandwidths:
1) 10 Hz - 3 kHz
2) 3 kHz - 100 kHz
3) 10 Hz - 100 kHz
Remark: the LCD indicates the selected
frequency bandwidth.
Range:
The choice of the range (200 or 2000 V/m)
is automatic.
Low bat:
is displayed when the battery is low.
Auto-off:
The appliance switches off automatically
after 30 minutes if no keys are used (the
appliance can be reactivated by pressing
one of the keys). A sound signal is given 2
secs. Before automatic switch-off.
Backlight:
Activated when the appliance is switched
on.
Automatic timed switch-off (15 secs).
Reactivate by pressing one of the keys.
Remark: for the
light is turned off.
Switching off the VX0100:
Press the "ON/OFF" key for approximately. 2 secs; the LCD
displays "OFF" and a high sound signal is emitted until the key
is released.
VX0003 / VX0100
Display: 0 - 3 kHz
Display: 3 - 100 kHz
Display: 0 - 100 kHz
the
measured
field
key.
function the back-
27

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Biotest vx0100

Table des Matières