Télécharger Imprimer la page

Specification - Amprobe TMD-55 Mode D'emploi

Masquer les pouces Voir aussi pour TMD-55:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Err-01: In the "SAVE" mode, if memory is full, the lower
right display will show Err-01. If you would like
to record the new value. Please clear all old
records.
Err-02: In the "LOG" mode, if memory is full, the lower
right display will show Err-02. If you would like
to record the new value. Please clear all old
records.
Err-03: In the "READ LOG" and "READ SAVE" mode,
when the main display shows 6208 and the
lower right display shows "OL, Err-03". Which
means that there might appear two situations:
1. There is no data in the memory.
2. The memory is full, and the meter will warn
the user this is the last sample.

SPECIFICATION

Electrical
Temperature Scale: Celsius or Fahrenheit user-selectable
Measurement Range:
Thermocouple
Type
K-TYPE (0.1°)
J-TYPE (0.1°)
T-TYPE (0.1°)
E-TYPE (0.1°)
R-TYPE (1°)
S-TYPE (1°)
N-TYPE (0.1°)
*Based on the ITS-90 temperature standard.
According to temperature standard ITS-90.
-200°C to 1372°C (-328°F to 2501°F)
-210°C to 1200°C (-346°F to 2192°F)
-200°C to 400°C (-328°F to 752°F)
-210°C to 1000°C (-346°F to 1832°F)
0°C to 1767°C (32°F to 3212°F)
0°C to 1767°C ( 32°F to 3212°F)
-50°C to 1300°C ( -58°F to 2372°F)
16
Range

Publicité

loading

Ce manuel est également adapté pour:

Tmd-55wTmd-56