Function And Limitations Of Use - BRAND macro Mode D'emploi

Masquer les pouces Voir aussi pour macro:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

The instrument is designed to assist the filling and dispensing of gradu-
ated and volumetric pipettes of glass or plastic in the volume range
of 0.1 ml to 200 ml with a suction tube outer diameter < 9.2 mm for
measuring liquids. If the instrument is used correctly, the pipetted liquid
will only contact the pipette.
Limitations of Use
The instrument is designed for pipetting liquids, observing
the following physical limits:
– from +10 °C to +40 °C (from 50 °F to 104 °F)
(instrument and reagent)
– Vapor pressure up to max. 500 mbar.
Aspirate slowly above 300 mbar, in order
to prevent the liquid from boiling.
– Density up to 2.4 g/cm
Operating Exclusions
Never use the instrument with liquids whose vapors
attack the materials silicone or PP.
The instrument is not designed for Pasteur pipettes.
Storage Conditions
Store the instrument and accessories only in cleaned
condition in a cool and dry place.
Storage temperature: - 20 °C to + 50 °C (-4 °F to +122 °F).

Function and Limitations of Use

3
21

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

262 00262 01262 02262 03Macro

Table des Matières