Télécharger Imprimer la page

Use; Operating Procedures; Powering On The Controller; Starting The Pump - Agilent Technologies Turbo-V 70 PCB 24 Vdc Notice De Mode D'emploi

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 34

Use

This section provides all the major operating procedures. For further
details and for the procedures that involve connections or specific
options, refer to the section entitled "USE" in the Appendix
"Technical Information".
Perform all electrical and pneumatic connections before using the
controller, with reference to the manual of the connected pump.
If the pump is placed on a table, make sure that it is stable so as to avoid
WARNING!
injury to persons and damage to the equipment. Never operate the pump
unless the input flange is connected to the system or closed using the closure
flange.

Operating Procedures

Powering on the Controller

To power on the controller, simply provide electrical power to it.

Starting the Pump

Per avviare la pompa lasciare aperto il pin 12 del connettore J4, in
modo tale che quando viene fornita la tensione di 24 Vdc al controller
si ha l'avviamento della pompa.

Stopping the Pump

To stop the pump, short-circuit pins 12 and 15 of connector J4.
Turbo-V 70 PCB 24 Vdc Controller User Manual / 87-900-933-01 (H)
Instructions for Use

Use

169/208
16

Publicité

loading

Ce manuel est également adapté pour:

969-9518