17
Technical Information
Data Exchange Protocols

Serial Communication Windows

Tab. 6
WIN
TYPE
R
000
L
X
001
L
X
008
L
X
100
L
X
101
N
X
102
N
X
103
N
X
104
L
X
105
N
X
107
L
X
108
N
X
109
L
111
N
X
120
N
X
121
N
X
196/208
W
Description
X
START/STOP (1= START ; 0= STOP)
X
SPEED SELECTION ACTIVE [0=HS / 1=LS]
X
REMOTE/SERIAL Configuration (1= Remote ; 0= Serial)
X
SOFT START YES/NO (1= YES ; 0= NO)
Default = 0
X
SET POINT TYPE (0 =Frequency; 1=Current; 2=Time)
Default = 0
X
SET POINT VALUE [Hz, Ma, SEC]
Default = 1000 Hz
X
SET POINT MASK TIME [s]
Default = 0
X
OUTPUT STATE (0= OFF if Val > threshold; 1 = ON if Val > threshold
Default = 0
X
HYSTERESIS IN THRESHOLD % [0-100]
Default = 2
X
ACTIVE STOP (0=NO; 1=YES)
Default = 1
X
BAUD RATE (3-4) [4800-9600]
Default = 4
X
PUMP LIFE RESET [Write "1" to Reset]
X
SET ANALOG OUTPUT TYPE [0=Power; 1=Current]
Default = 0
X
SET ROTATIONAL FREQUENCY [Hz] 150 Hz < = F_imp < = FMAX
X
MAX SETTABLE ROTATIONAL FREQUENCY [Hz] F<=F_MAX_ABS
Turbo-V 70 PCB 24 Vdc Controller User Manual / 87-900-933-01
(H)

Publicité

loading

Ce manuel est également adapté pour:

969-9518