Problem
Lamp does not
•
go on.
•
•
•
Compressed air
•
leaking from the
foot switch.
Foot switch fails
•
to switch on.
•
Foot switch hoses
•
cannot be discon-
nected from the
unit.
EN
Possible cause
Power cord not plugged in.
Fluorescent tube defective.
Faulty electronics.
Fluorescent tube not prop-
erly seated in the socket.
Pneumatic hoses re-
versed.
No pressure at the com-
pressed air connection.
Pneumatic hoses are
kinked.
Push-in connection is not
released (Fig. 1).
Corrective action
•
Plug the power cord into the wall outlet.
•
Replace the fluorescent tube (refer to Sec. 5.6).
•
Call for service to replace the electronics, or refer to the spare
parts list.
•
Refer to Sec. 5.6, "Fluorescent Lamp Installation".
•
Reverse the connections.
•
Inspect the compressed air lines and/or the compressor.
•
Reposition the hoses.
•
Press the push-in connection ring on the hose side firmly
while pulling on the hoses.
- 18 -