Structure and mode of operation
14
3.4.2
Filling and dosing
The electronics include full dosing control (command
IMD2). As many as 32 parameter sets can be stored in
the EEPROM for different applications. But you can still
change the dosing parameters yourself during dosing.
Digital outputs can be used to control coarse and fine
flow, for example. The PanelX software includes detailed
instructions for setting the different parameters.
3.4.3
Limit value function
In Standard and Trigger modes (command IMD), the
electronics allow as many as four limit values to be mon
itored (command LIV). Gross or net values, the trigger
result, or the extreme values (Min/Max) are available to
you as input signals. Use the measurement status to
read out the status, either simultaneously with the meas
ured values (command MSV?) or separately (command
RIO?).
3.4.4
Extreme value functions
The electronics include a peak value function (Minimum
and Maximum, command PVS), that monitors either the
gross or net values, as required. Use command PVA to
read out the values and use command CPV to reset the
peak values.