Adaptive Interference Suppression - HBM PAD4001A Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Structure and mode of operation
Output format
Binary, 2 chars. (INT)
Binary, 4 chars. (LONG)
ASCII
12
S activate an automatic zero on start up function,
S activate an automatic zero tracking function,
S linearize the input signal with a third order polynomial,
S activate various digital filters. Available filters include
those with cut-off frequencies below 1 Hz, fast-settling
filters for dynamic measurements, notch filters and
mean value filters.
Use MSV? to read out the current measured value. The
format of the measured value (ASCII or binary) is set
with the COF command. You can also use the COF com­
mand to activate automatic measurement output. The
measured values are transmitted in the following format,
subject to the COF command:
Input signal
0 ... maximum
capacity
0 ... maximum
capacity
0 ... maximum
capacity
3.3

Adaptive interference suppression

Whatever the mode of operation, you can use the ADF
command to activate automatic interference suppression
with adaptive filters. Interference frequencies are auto­
matically found during measurement and suppressed by
comb filters and averaging. The maximum filter settling
time can be limited with the TMA command.
A4238-1.0 HBM: public
Output when NOV=0
0 ... 20,000 digits
0 ... 5,120,000 digits
0 ... 1,000,000 digits
Output when
NOV>0
0 ... NOV
0 ... NOV
0 ... NOV
PAD4001A/4003A