The data output can be requested from the computer by toggling the DTR line in the following modes:
See also user manual of OPTO-RS.
VIII.5 Use with the half duplex adapter
With the half duplex adapter, you may completely control the instrument from the computer by sending remote commands in the following modes
See OPTO-RS user manual for connexion on a computer
VIII.5.1 Remote command names
Most commands consist of 3 characters followed by either a 0 or 1 (disabled or enabled). Note: commands may be upper or lower case and should be followed by
the ASCII code <CR>.
The instrument echoes the command.
VIII.5.2 List of remote commands
<NOR>
Will set the instrument in the measure mode (or in reference mode if the button is disabled)
<MOD?>
The instrument returns its current mode of operation (NOR, REF, MIN,MAX, DEL, TOL1)
<STO0>, <STO1>
Deactivates or activates the measurement "HOLD"
<KEY0>, <KEY1>
Deactivates or activates [mode] button
<RST>
Reset the instrument to the factory default condition
<SET?>
The instrument returns its current parameters (MM RES2 REF1 etc)
Note:
- B1 Battery OK
- B0 Battery must be replaced
<ID?>
The instrument returns its class identification
- SY210A: instrument basic
- SY210B: instrument complete with preset, minmax, tol. mode.
<OUT0>, <OUT1>
Deactivates or activates continuous data output.
<OFF>
Turns off the instrument
<ON>
Turns on the instrument
<PRI>, <?>
The instrument returns the displayed value.
Remark: if tolerances mode is active, ASCII code '<', '=' or '>' is added.
<CHA+>, <CHA->
Changes the measuring direction
<CHA?>
The instrument returns the active measuring direction.
<CHA*1>, <CHA*2>
Changes the multiplication factor
<MM>, <IN>
Changes the measurement unit
<RES2>, <RES3>
Changes the resolution:
-<RES2>: 0.001 mm
.00005 "
-<RES3>: 0.01 mm
0005 "
<REF1>, <REF2>
Changes the reference
<PRE>
Recalls the last preset value
<PRE?>
The instrument returns the last preset value
<TOL1>
Activates the tolerance mode
<TOL?>
The instrument returns the tolerance limit values
<MIN>, <MAX>, <DEL>
DEL = Delta = [MAX-MIN] Selection of the dynamic mode
<CLE>
Resets the MIN, MAX memory
<PRE + 123.45>
<PRE + 0>
Presets the display to the required value.
Note: don't forget to use the correct sign preceding the value
- measuring
- references
- minmax
- tolerances
- measuring
- references
- minmax
- tolerances
65
65
65
65