Unpacking
• Unpacking
- Unpack the device carefully
- Any damage should be notified immediately to the shipping
agent (post office, railway network or transport company)

Correct use

• Use
- for mixing and/or heating liquids.
• Area of use (only indoors)
- laboratories
- pharmacies
The device is suitable for use in all areas apart from domestic areas
and areas directly connected to a low-voltage supply which also
serves domestic areas.

Assembling the support rod

• R emove the threaded plug (P)
• R emove the protective cap from the support rod
• P lace the washer between the housing and the nut
• S crew in the support rod manually until it cannot be
tightened further
• T ighten the nut using an SW17 open-end spanner
• A ssemble the accessories using a boss head clamp

Setting the safety temperature limit

The set safety temperature is displayed permanently (7).
Factory settings: approx. 380°C
Range: 50 - 380°C
Setting the safety temperature limit
a S et the main switch (A) to the ON position
a S et the temperature limit using the screwdriver (supplied)
at point (B)
i The set value is displayed on the screen
The adjustable maximum heating plate temperature of 340°C is
at least 10°C under the set safety temperature limit.
Caution: the set safety temperature limit must always be
at least 25°C below the flashpoint of the medium in use.
- schools
- universities
• Scope of delivery
- RET control visc heated magnetic stirrer
- Power supply cable
- Operating instructions
- Screwdriver (safety circuit)
- USB cable
- Protective cover: H104
- Magnetic rods: IKAFLON 30 and 40mm
- Temperature sensor: PT 100.70
The safety of the user cannot be guaranteed if the device is used
in conjunction with accessories which are not supplied or recom-
mended by the manufacturer, if the device is used contrary to the
intended purpose against the manufacturer's instructions or if the
device itself or PCB has been subjected to modifications by third
parties.
Note:
When using bath attachments with a diameter
over 180 mm, use sup-
port rod H 16 V in con-
junction with an  H
16.1 extension.
Functional test: safety circuit switch-off
a H eat the device to a temperature of over 50°C.
a S et the safety temperature limit as far to the left as possible (50°C)
and set the main switch (A) to the ON position
i The display will show:
Be careful not to turn the adjustment screw too far to the left or
right or this will damage the potentiometer.
Err 25
22