Before loading the machine and before starting a
programme
– Check that fittings such as irrigation sleeves and injector nozzles
are securely screwed into position.
Make sure that all screw connectors are fitted with injector
nozzles, adapters, irrigation sleeves or blind stoppers to ensure
that all fittings in use are supplied with sufficient standardised
pressure.
Damaged fittings such as injector nozzles, adapters and irrigation
sleeves must not be used.
Fittings not equipped with wash items do not have to be replaced
blind stoppers.
– Check that the baskets and modules are correctly connected to the
water supply.
– Make sure that the measuring cylinders are not touching each other
or resting against the wash cabinet walls.
Reprocessing
Slowly and carefully insert the lower basket with the loaded module
into the cleaning machine.
Make sure that the supporting feet of the measuring cylinders do
not stick out into the door opening or the door seal or collide with
the wash cabinet walls.
Carefully close the cabinet door.
Protruding measuring cylinders which extend into the door area
may be damaged when the door is closed or during reprocessing.
en - Areas of application