Connect NIOX MINO
using USB
Caution!
• A PC connected to the USB connector
has to be certified for one of the
standards IEC 60601-1, IEC 61010-1, IEC
60950 or comparable with safety extra
low voltage on the USB ports.
• The connected PC should be placed
out of reach from the patient. Do not,
simultaneously, touch the connected PC
and the patient.
1. Connect the USB cable from the instrument to
the PC
2. The USB icon
display to indicate that a USB connection is
established
to a PC
®
USB cable from PC
is shown on the NIOX MINO
Connect NIOX MINO to a PC
using Bluetooth
1. Select Settings
2. Select Bluetooth
3. The Bluetooth setting
is shown
4. Use the button to
activate Bluetooth
5. Select OK to accept
the changes
6. Follow the instructions delivered with the PC or
from the Bluetooth device manufacturer for how
to set up a Blutetooth connection:
• Create a new Bluetooth connection
• Select NIOX MINO
• Enter passcode 0000
E N G L I S H