E N G L I S H
Information menu
1. Select Settings
2. Select Information menu
3. The information screen shows:
A. Number of remaining sensor
measurements
B. Sensor expiration date
C. Sensor serial No.
D. Number of remaining
instrument measurements
E. Instrument expiration date
F. Instrument serial No.
G. Instrument software version
H. Ambient RH (relative
humidity), measured inside the
instrument (the value may differ
from RH measured outside the
instrument)
4. Select Return to go back to the
Settings screen
... alternatively select Alert to view
the alert codes
5. The alert code information screen
is shown, showing the 16 latest
alerts
Note!
The alert codes in the list are
only to be used in contact with
Customer support
• Date of the alert
• Alert code (for customer support
purpose only)
6. Select Return to go back to the
Information screen
10
Turn off NIOX MINO
Note!
When within the specified operating
conditions (see page 16) it is recommended
to keep NIOX MINO with installed Sensor
connected to a power outlet whenever
possible.
A
1. Pull out the power supply unit connector from
B
NIOX MINO
C
Moving NIOX MINO
D
The instrument can be moved to another power
E
outlet with optimized initiating startup.
F
Note!
G
The instrument must be connected to the new
H
power outlet within a few minutes.
1. Select Settings
2. Select Controlled power
off
3. Select OK to confirm
4. Move the instrument and
connect it to a power
outlet
5. Wait until the instrument
is ready to use
®