General care
• Use a new NIOX
Filter (patient filter) for each
®
new patient.
• Clean the instrument with a cloth dampened with
70% ethanol, isopropanol or similar disinfectant.
Be careful when using disinfectants, as excess of
alcohol (ethanol) might permanently damage the
Sensor. Do not use spray detergents.
Note!
Never attempt to open or service the
NIOX MINO
instrument or Sensor.
®
Preventive inspections
Before each measurement, visually inspect:
• that NIOX MINO is not damaged
• that the Sensor is in place
• that a NIOX Filter is attached
• that the Power supply cord is undamaged and
correctly connected to the instrument and to the
power outlet.
If any item is missing or damaged, contact your
local Aerocrine representative or Aerocrine AB.
Change Sensor
Caution!
Be careful when opening the Sensor can.
The inside of the opening has sharp edges.
Do not touch or clean the white Sensor
membrane
1. Press and hold the blue
button...
...while turning the orange
swivel
2. Remove the Sensor
3. Insert the new Sensor
4. Turn back the orange
swivel until locked
Change NO scrubber
1. Spread the latches apart,
one at a time,
(Optionally: Use the red
tool delivered with the
instrument to spread the
latches apart)
... and carefully pull out
the cover
2. Remove the used
NO scrubber from the
cover
3. Insert the new
NO scrubber into the
cover
Make sure to use a new
NO scrubber, with an
unbroken hole
4. Replace the lid and make
sure it snaps in place
5. Select Settings
6. Select NO scrubber
7. Again select NO scrubber
E N G L I S H
Unbroken
Broken