E N G L I S H
Symbols explanation
Responsible manufacturer
The product meets the requirements of
applicable European directives
ETL-Listed. Conforms to ANSI/AAMI
ES60601-1 and certified to CAN/CSA
C22.2 No. 60601-1.
Electrical safety Type B applied part
100-240 V ~50-60 Hz
The product should be recycled
according to the local program for
electronic equipment.
Consult instructions for use
Refer to instruction manual/booklet
Expiration date
Temperature limitation
For single use only
In vitro diagnostic medical device
Durability - period after opening
12M
Hazardous material
Class II equipment
Backplate
1
2
3
4
5
1. Responsible manufacturer
2. Refer to instruction manual/booklet
3. In vitro diagnostic medical device
4. Type B applied part
5. The product should be recycled according to the
local program for electronic equipment
6. The product meets the requirements of
applicable European directives
7. ETL-Listed. Conforms to ANSI/AAMI ES60601-1
and certified to CAN/CSA C22.2 No. 60601-1
16
6
7
Base label
1
2
3
1. Instrument expiration date (year:month)
2. Instrument article number
3. Instrument serial number
Dimensions and weight
Height: 230 mm
Width: 128 mm
Depth: 96 mm
Weight of instrument including Sensor: 0.8 kg
Electrical data
Electrical safety classification: The equipment
complies with the requirements according to IEC
61010-1 and 60601-1 Class II Type B.
NIOX MINO instrument
Secondary voltage (external power adapter): 6 V
Power consumption: < 15 VA
Power supply unit
Mains Voltage: 100-240 V ~50-60 Hz
Noise level (stand by)
< 35 dBA, at a distance of 1 m
Shelf-life
NIOX MINO
instrument: Minimum 3 years at time
®
of delivery, or 3,000 measurements.
NIOX MINO Sensor: Maximum 12 months when
installed in NIOX MINO or expiration date as stated
on the Sensor, whichever comes first.
Always use the NIOX MINO Bag (closed) for
transportation and storage of NIOX MINO
when it is disconnected from the power outlet.
(The NIOX MINO Bag is sold separately).
Operating conditions
Ambient temperature: +16 to +30°C
Humidity: 20 to 60% RH (non-condensing)*
Atmospheric pressure range: 700 to 1060 hPa
Ambient NO concentration: <300 ppb
Exhaled flow rate during measurements at 50 ml/s
± 10% BTPS (Body Temperature Pressure Saturated,
37˚C, 1013 hPa, 100% RH)
* The ambient RH in the instrument can be viewed
in the information menu, see page 10 and may
differ from RH measured outside the instrument.