You can order suitable alkaline batteries stating the following item no.:
item no. 65 25 09 (please order one).
Only use alkaline batteries, as they are powerful and have a long service life.
d) Fuse Change
The current measuring ranges are protected against overload with high-performance fuses. If no measurements in
the current measuring range are possible anymore, the fuses are probably defective and need to be replaced.
Always observe the safety provisions during use changes!
Ensure that only fuses of the indicated type and rated current must be used as spares. Use of
wrong or patched-up fuses or bridging of the fuse holder is not permitted and may cause fire.
Proceed as follows for fuse replacement:
- Disconnect the connected measuring lines from the measuring circuit and the meter. Switch off the DMM.
- Loosen the four rear housing screws with a
matching Phillips screwdriver. Carefully loosen the
two housing halves from each other and turn away
the rear as shown. Observe the battery cable
- Replace the defective fuse with a new fuse of the
same type and rated current.
FUSE1: FF200 mA 600 V
FUSE2: F10A 600 V
However, always observe the information at the
device or the values of the fuse used.
- Close the casing carefully again.
Never operate the meter when it is open. !DANGER TO LIFE!
40
5 mm x 20 mm