Water is the only coolant that may be used (with antifreeze, e.g. ethylene
glycol).
Permissible cooling agents - inlet temperature >3°C.
Observe the maximum permissible pressure of 1 bar! As a safeguard, we rec-
ommend using a pressure limiter (e.g. IKA C25). This is not necessary when
using an IKA KV600.
Accumulated condensate is drained out of the cooler through the drain hose.
Attachments
The following attachments are offered as accessories:
AS 4000.1 universal attachment
AS 4000.3 bowl attachment
The attachments are affixed to the shaking table using the four clamping screws
provided.
Load (standard values)
Load / kg
6
Max. speed / rpm
500
AS 4000.2 bracket attachment
12
15
20
400
300
120
Assembly information for AS 4000.1
Error codes
An error is indicated by an acoustic signal and an error code is displayed.
E.g.:
Note:
During continuous operation, the load may slip, for example, resulting in the appliance
becoming excessively unbalanced. To prevent this, there is a motion sensor built in
which reduces the speed gradually once a critical limit has been reached.
When faults are not specified in the table, you switched the appliance off and again on.
If faults cannot be eliminated directly, you must perform a RESET (see "Reset")! If
the faults still cannot be eliminated, the appliance must be inspected by a technical
service.
21