6.1 Cleaning
If the instrument is dirty after usage, we advise to
clean it by using a humid cloth and mild household
detergents. Never use acid detergents or dis-
solvants.
6.2 Fuse replacement
Fuse replacement High Voltage Tester contains two
built-in fuses. If a fuse triggers, return the instru-
ment for repair to our factory.
Fuse type
F1: 3,15A / 250V slow blow (5 x 20mm)
F2: 200mA / 250V slow blow (5 x 20mm)
6.3 Calibration interval
Measurement instruments are to be calibrated in
regular intervals, in order to guarantee accuracy and
reliability of measurement results. A calibration in-
terval of one year is recommended for High Voltage
Tester 0113.
7.0 Technical Data
Display:
LED Thermal Overload,
Insulation Breakdown,
Power on, Stand By and
TEST
Test voltage:
1000V, 1500V AC
(at nominal load)
Max. Test current:
500 mA (at 1000 V), 333 mA
(at 1500 V)
Temperature ranges: 0° C...+40° C
(operating temperature)
-10° C...+60° C
(storage temperature)
Hight above MSL:
up to 2000 m
Excess temperature protection:
Safety:
Class II,
DIN VDE 0411, Part 1
EN 61010, IEC 61010-1
DIN VDE 0404, Part 1
Overvoltage category: CAT II / 250 V
Pollution degree:
Power supply:
approx.
Dimensions:
Weight:
12 months warranty
UNITEST instruments are subject to strict quality
control
However, should the instrument function improp-
erly during normal use, you are protected by our 12
month warranty (valid only with invoice or receipt).
Within the warranty period we will decide whether
to exchange or repair the defective instrument. We
will repair free of charge any defects in workman-
ship or materials, provided the instrument is re-
turned unopened and untampered with.
Damage due to dropping or incorrect handling are
not covered by the warranty. If the instrument
shows failure following expiry of warranty our ser-
vice department can offer you a quick and econom-
ical repair,
for +60° C
Subjects to change without notice
Maintenance / Technical Data
2
230V 10% 50-60Hz
500VA
250x160x145 mm
(LxWxH)
approx. 7 kg
15