Synchronisation (For Hls 528-2 And Hls 528-3 Only) - HYDAC ELECTRONIC HLS 528 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

HLS 528
Important notes
In order to get precise measuring results of the distance to an object/surface, it should be
ensured that no interfering objects project into the blind zone or into the detection range.
These will produce interference reflections which will distort the measurements.
Minimum distances to be kept in order to avoid interference reflections:
Detected objects
Min pipe diameter 10 mm
Plate, fluid surface
Detected objects
Min pipe diameter 27 mm
Plate, fluid surface
3.1

Synchronisation (for HLS 528-2 and HLS 528-3 only)

If the assembly distances shown in Fig.1 for two or more sensors are exceeded the integrated
synchronisation should be used. Connect Sync/Com-channels (pin 5 at the units receptable) of
all sensors (10 maximum).
HLS 528 - X - 0250
HLS 528 - X - 0350
HLS 528 - X - 1300
HLS 528 - X - 3400
HLS 528 - X - 6000
Fig. 1: Assembly distances, indicating synchronisation/multiplex
Status 12.06.2014
Sensor
Blind zone
Minimum distances
Recording ranges
Cylinder
Plate, fluid surface
Operating scanning range
250 mm
3 cm
5 cm
Operating scanning range
1300 mm
3400 mm
15 cm
40 cm
35 cm
90 cm
HYDAC ELECTRONIC GMBH
350 mm
6 cm
10 cm
6000 mm
50 cm
130 cm
Page 6
Part No. 669826 E