Therapy time can be increased up to 23 hours and 55 minutes. It's
possible to quickly scroll time list pressing TM. key;
6)
MAG1000 is now ready for use: the display shows information about
the program in use, the working frequency (not adjustable), the
adjusted therapy time and field intensity (adjustable during the
treatment);
7) Press
while it will be possible to modify field intensity during the whole
treatment, pressing more times IN. key. The relative icon is positioned
on the lower right side of the display, with the indication of the three
intensities (1 - low intensity,2 - medium intensity-two bars and 3 - high
intensity);
8) The therapy can be suspended pressing the key
will show a flashing icon
therapy press again
9) To end the therapy press key
the selected program and adjusted therapy time. To switch off
MAG1000, press again key
10) A sound signal inform of the end of the program, and the display will
show the flashing icons 0'00''; to start again a new therapy , press
key
and then follow again the indications from step 4);
WARNING: the display shows two icons relative to channel A and B. The
flashing icon shows the not in use channel (no therapy belt connected).
WARNING: MAG 1000 will enter into "sleep mode" if not press any key for
2 minutes to preserve batteries.
IACER S.r.l.
key : the display will show the remaining therapy time,
;
34 di 56
on the lower right side. To restart
: MAG1000 will shown again
;
MNPG88 Rev. 00 Ed. 03/05/2012
: the display