CLEANING AND DISINFECTING

Cleaning
1. Remove all accessories (including nebulizer, mask, mouthpiece,
connector, outer cover) from the device after each use.
2. Open the cover of nebulizer and remove the residual liquid. You are
advised to, after each use, allow the device running with clean water
instead of liquid for 1-2 minutes in order to clean the nebulizing head.
Clean the mask, mouthpiece, connector, outer cover, nebulizer gently
(Note that the main unit should not be washed).
3. Immediately after completion of cleaning, dry the accessories with
medical gauze. Note that the metal mesh in nebulizing head must not be
touched strongly any sharp and rough object. Dry the water inlet
and outlet of the nebulizing head with medial swab gently.
4. If there is dust on the surface of main unit, please wipe it up with wet
and dry gauge successively.
5. After completion of cleaning, store the device in clean storage bag.
Disinfecting
1. Rinse the mask, mouthpiece, connector and outer cover with solution
of ammonium chloride diluted to 0.1% for 10 minutes (Not including
the MESH HEAD); rinse with water; dry with medical gauze.
2. Disinfecting of nebulizing head: gently clean the metal mesh and
electrode in nebulizing head with medical swab as dipped with medial
alcohol; rinse the nebulizing head with water; dry the surface of
nebulizing head with medial swab as soon as possible by patting action.
3. Boiling Disinfection for the following accessories:
Mouthpiece, connector and outer cover: put them in boiling sterilizer with
sufficient water to ensure they can be immerged completely; heating the
water to boiling point (for 20 minutes in total); dry them with medial
gauze.
4. Boiling Disinfection must NOT be used for the following parts:
Nebulizer, nebulizing head, main unit, mask, power adapter, storage bag,
packaging bag.

STORAGE AND MAINTENANCE

• Make sure no residual liquid in nebulizer.
• Make sure the nebulizer has been cleaned and sterilized thoroughly after
each use.
• Store the device and its parts and components and accessories in clean
and safe place.
• Protect the device against direct sunshine and shock.