Error messages
Error
Possible cause
description
Power pack or battery voltage
Err 5
too low.
Battery contacts are corroded.
Err 6
Air congestion
Blood pressure cuff incorrectly
connected.
Leak in the blood pressure cuff
or air hose
Err 9
Residual pressure in the blood
pressure cuff
Err 10
Internal error
Abr. (Abort)
Measurement cancelled on
request
44
Remedy
Change the batteries (Chapter 5.2).
Clean the battery contacts with a cotton cloth and a little
alcohol.
Check the blood pressure cuff for air congestion or a kink in
the air hose. If there is a kink in the air hose, unkink it,
otherwise send the blood pressure monitor to your
healthcare provider/physician or the manufacturer.
Connect the blood pressure cuff to the blood pressure
monitor
(Chapter 6.2).
Replace the blood pressure cuff with the air hose.
Wait until the blood pressure cuff has deflated completely.
If this error occurs repeatedly, send the blood pressure
monitor to your healthcare provider/physician or directly to
the manufacturer for checking.
Do not press this button when measuring, unless you want
to abort the measuring.