Télécharger Imprimer la page

Vaisala Polaris PR53AC Guide D'installation page 33

Publicité

Property
External connectors
1)
For USB2 adapter and Insight software, see www.vaisala.com/insight.
2)
Conduit hub is not compatible with SDI5 Safe-Drive system.
Environmental specifications
Table 7 PR53AC operating environment
Property
Process parameters
Process temperature
Design temperature
Design pressure
Operating environment
Storage temperature
Operating temperature
Maximum operating altitude
Operating humidity
Storage humidity
UL 50E/NEMA rating
IP rating
1)
–40 ... +130 °C (-40 ... 266 °F) EPDM gasket, -40... +150 °C (–40 ... +302 °F) PTFE gasket.
2)
Maximum momentary temperature peak.
3)
Maximum at +20 °C (68 °F), operating pressure to the clamp rating pressure.
Specification
1 × M12 M 4 pins, A-coded
2 × M16×1.5 cable gland, Cable D 5–10 mm /
Adapter for conduit entry M16×1.5
Specification
–40 ... +150 °C (–40 ... +302 °F)
2)
+180 °C (356 °F)
3)
40 bar
−40 ... +65 °C (−40 ... +149 °F)
−40 ... +60 °C (−40 ... +140 °F)
2000 m (approx. 6500 ft)
0–100 %RH
0–100 %RH, non‑condensing
Type 4X: Dust-tight. Protected from corrosion
and hose-directed water.
IP66: Dust-tight. Protected from powerful water
jets from any direction.
IP67: Dust-tight. Protected from the effects of
temporary immersion in water under
standardized conditions of pressure and time.
1)
2)
/ NPT ½"
1)