8 SPARE PARTS AND ACCESSORIES
8.1 Included With Your Unit
PART NUMBER ITEM
R221M11
Operating Manual and Instructions for Use*
R100P91-001
Gas Sample Tubing
R221P15
Soft Cover
Repair of this equipment must be performed by a qualified service technician experienced in
repair of portable hand held medical equipment.
Equipment in need of repair shall be sent to:
Maxtec
Customer Service Department
2305 South 1070 West
Salt Lake City, Ut 84119
(Include RMA number issued by Customer Service)
9 ELECTROMAGNETIC
COMPATIBILITY
The UltraMaxO2 is suitable for the electromagnetic environment of typical hospital and home
healthcare settings. The user should assure that it is used in such an environment.
During the immunity testing described below, the UltraMaxO2 will analyze oxygen concentration
within specification.
WARNING: Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the UltraMaxO2, including cables specified by the manufacturer. Otherwise, degradation
of the performance of this equipment could result.
WARNING: The UltraMaxO2 should not be used adjacent to or stacked with other equip-
ment. If adjacent or stacked use is necessary, the UltraMaxO2 should be observed to verify
normal operation. If operation is not normal, the UltraMaxO2 or the equipment should be
moved.
WARNING: Use of accessories, transducers and cables other than those specified or pro-
vided by the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in improper
operation.
WARNING: Avoid exposure to known sources EMI (electromagnetic interference) such as
diathermy, lithotripsy, electrocautery, RFI (Radio Frequency Identification), and electromag-
netic security systems such as anti-theft/electronic article surveillance systems, metal de-
tectors. Note that the presence RFID devices may not be obvious. If such interference is
suspected, reposition the equipment, if possible, to maximize distances.
WWW.MAXTEC.COM • (800) 748-5355
This equipment is intended for use in the electromagnetic environment specified below.
The user of this equipment should assure that it is used in such an environment.
EXPECTED SERVICE LIFE
EMISSIONS
N/A
Life of UltraMaxO2
RF Emissions
(CISPR 11)
Life of UltraMaxO2
CISPR Emissions
Classification
Harmonic Emissions
(IEC 61000-3-2)
Voltage Fluctuations
(IEC 61000-3-3)
The UltraMax O2 was also tested for radiated immunity to RF wireless communication equip-
ment at the test levels below
Frequency (HZ)
385
450
710, 745, 780
810, 870, 930
1720, 1845, 1970
2450
5240, 5500, 5785
7
ELECTROMAGNETIC EMISSIONS
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT
ACCORDING TO
Group 1
The UltraMaxO2 uses RF energy only for its inter-
nal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
Class B
The UltraMaxO2 is suitable for use in environments
of typical hospital and home healthcare settings.
N/A
N/A
Modulation
PULSE, 18 Hz, 50% DC
FM, 1 kHz Sine, ±5 Hz Deviation
PULSE, 217 Hz, 50% DC
PULSE, 18 Hz, 50% DC
PULSE, 217 Hz, 50% DC
Level V/m
27
28
9
28
28
28
9