• Insert the protrusion of the battery into the groove on the
back of the motor handpiece, and then push it until it clicks.
• To remove the battery, press the
at the bottom of the back of the motor handpiece.
WARNINGS
• Only use the battery, charger, AC/DC adapter, and power cord
included in the package.
• Be sure to plug the power cord into an outlet with a protective
grounding.
PRECAUTIONS
• If only one slot of the battery icon remains on or the icon flashes,
replace it with a charged battery and charge the used battery.
INITIAL SET-UP
• With the motor handpiece powered off, attach, and remove
the contra angle, Ni-Ti file, from the motor handpiece.
1
Assembling the Device
1. Insert the pin of the contra angle connector into the groove of
the motor handpiece connector and then push it until it clicks.
2. The contra angle rotates 360° and can be adjusted to work in
the desired position.
PRECAUTIONS
• It is recommended to use a plastic barrier with your handpiece to
ensure longevity of the device and prevent cross-contamination
between patients.
• If there is any residual oil on the surface of contra angle, wipe it
with the tissue and then alcohol gauze.
5
[Battery Release]
button
PRECAUTIONS
• Do not install any contra angle other than the contra angle
included in the package to the motor handpiece.
• When using the device for the first time or after long-term
storage, refer to "Cleaning, Disinfection, Lubrication,
Packaging and Sterilization Instructions" to sterilize
the device.
• The contra angle is not sold sterile and must be sterilized prior
to first use.
3. While holding down the
insert the Ni-Ti file into the contra angle latch. Then turn it
slightly to the left or right, and slide it into the latch groove.
4. Pull the inserted Ni-Ti file gently to make sure it is
properly secured.
PRECAUTIONS
• If the Ni-Ti file is not properly secured to the contra angle, the file
may come off during use and may cause injury to the patient.
• Do not use deformed or damaged Ni-Ti files.
2
Disassembling the Device
1. While holding down the
pull out the Ni-Ti file.
PRECAUTION
After using the device, refer to "Cleaning, Disinfection,
Lubrication, Packaging and Sterilization Instructions"
to sterilize the contra-angle and to disinfect the motor.
3
Turning the Device On/Off
1. Press the
[Main]
button on the motor handpiece. The power
turns on, and the main screen appears on the display.
2. To turn off, press any two buttons simultaneously for 2 seconds.
CAUTION
• After using the device, refer to "Cleaning, Disinfection,
Lubrication, Packaging and Sterilization" instructions to
sterilize the contra angle and to disinfect the motor.
[Push]
button of the contra angle,
[Push]
button of the contra angle,