Télécharger Imprimer la page

KERR ENDODONTICS Elements Connect Mode D'emploi page 4

Publicité

CHARGING BASE
No.
Name
1
Device stand
2
Charging indicator
3
Battery charging
unit
4
Power connector
terminal
LCD SCREEN
The main screen is the screen that appears on the display when
the motor handpiece is turned on. The memory number and
motor operation information can be viewed and set.
Figure 1 - Main screen
No.
1
Memory number.
2
[EAL]
appears when the Apex Connect and motor handpiece
are connected properly.
3
This is the battery icon. The remaining battery is displayed.
4
The motor operation information is shown here. Motor rotation,
speed, torque limit, rotation angle and autoreverse options
are displayed.
Motor
CW
Rotation
CCW
Settings
Adaptive
RCP
Function
Stores the motor handpiece.
Displays the charging status of the
battery. (Charging: Orange, Charged: Blue)
Charges the battery.
Connects the power cord and AC/DC
adapter to supply power to the charger.
Description
Motor Settings
Motor rotates clockwise
Motor rotates counterclockwise
• Adaptive Motion relies on a patented
feedback algorithm that changes the
motion of the file based on the applied
load (stress). When the file is doing
minimal work—when there is no stress
on the file—the motion is purely rotary
(clockwise rotation only).
• When the file begins to cut more
dentin, the motion begins to "adapt"
from pure rotary to reciprocation
(clockwise and counterclockwise
rotation). This action disengages the
file from grabbing too much of the
canal wall.
• The Elements Motor adapts the motion
based on the amount of pressure on
the file. It may or may not reciprocate
and does not pull into the canal.
Reciprocating angles vary: no load 600°
forward and 0° backwards, loaded 370°
forward and up to 50° backwards, based
on file load.
The motor alternates between
clockwise and counterclockwise rotation
(Reciprocation).
Auto-
R&S
reverse
Settings
REV
Motor
RPM
Speed
Motor
Torque
Torque
Rotation
Degrees
Angle
(in the RCP
mode)
System
Dominant
Settings
hand
Volume
Autopower
off
Apex
EAL
Locator
Auto Start
& Stop
Apical
Action
Default Settings
Memory
Name
RPM
Code
TRVO
Kerr Traverse
500
Opener
TRVG
Kerr Traverse
500
Glide Path
ZF 04
ZenFlex 0.04
500
Small
20-25
ZF 06
ZenFlex 0.06
500
Small
20-25
ZF 04
ZenFlex 0.04
500
Large
30-45
ZF 06
ZenFlex 0.06
500
Large
30-45
K3
K3 / K3XF
350
ADP
Adaptive
N/A
TF
TF files
500
RCP
Left Handed
N/A
Reciprocation
Motor Settings
When the set torque limit is reached, the
motor rotates in the opposite direction of
the current setting until the load decreases
and stops.
When the torque limit value is reached, the
motor rotates in the opposite direction of
the current setting until the load decreases.
It then rotates again in the currently set
direction.
100-1000
1 to 4.0 Ncm/100 to 400 gcm
20° to 360°
The device window can be changed from
right hand to left hand use. (Default: Right)
VOL.0 to VOL.3. (Default: VOL.2)
Standby time can be set between 1 and
10 minutes, in 1 minute increments.
(Default: 5 min)
Visual indicator that the motor is correctly
linked to the Apex Connect.
Operates the motor:
ON: The motor operates when the file is
inserted into the root canal, and the motor
stops when the file is pulled out from the
root canal.
OFF: The motor does not run even if the file
is inserted into the root canal.
When the Apex Locator measurement
approaches the APEX reference value during
root canal expansion, the speed of motor
and torque of motor decrease so that the
root canal does not expand deeper than
the set value.
Speed Slowdown: When the Apex Locator
measurement approaches the APEX
reference value, the motor's rotational
speed decreases.
Torque Slowdown: When the Apex Locator
measurement approaches the APEX
reference value, the motor's torque limit
decreases.
OFF: The motor operates the same
regardless of whether or not the Apex
Locator measurement approaches the APEX
reference value.
Torque
Motor
Auto-
Rotation
reverse
350 gcm
CW
(3.50Ncm)
150gcm
CW
(1.50 Ncm)
100gcm
CW
(1.0 Ncm)
300gcm
CW
(3.0 Ncm)
200gcm
CW
(2.0 Ncm)
350 gcm
CW
(3.5Ncm)
300gcm
CW
(3.0Ncm)
N/A
ADP
400gcm
CW
(4.0 Ncm)
N/A
CCW 150
CW 30
REV
REV
REV
REV
REV
REV
REV
REV
REV
REV
2

Publicité