Property
Digital outputs
Digital output
Maximum cable run
Supported protocol
Connectors
External connectors
1)
For USB2 adapter and Insight software, see www.vaisala.com/insight.
2)
Conduit hub is not compatible with PR53 Safe-Drive system
Environmental specifications
Table 6 PR53GC operating environment
Property
Process parameters
Process temperature
Design temperature
Operating environment
Storage temperature
Operating temperature
Maximum operating altitude
Operating humidity
Storage humidity
UL 50E (NEMA) rating
IP rating
1)
Maximum momentary temperature peak.
28
Specification
RS-485, non-isolated
300 m (approx. 1000 ft) (digital)
Modbus RTU
1 × M12 M 4 pins, A-coded
2 × M16×1.5 cable gland, Cable D 5–10 mm /
Adapter for conduit entry M16×1.5
Specification
–40 ... +150 °C (–40 ... +302 °F)
1)
+180 °C (356 °F)
−40 ... +65 °C (−40 ... +149 °F)
−40 ... +60 °C (−40 ... +140 °F)
2000 m (approx. 6500 ft)
0–100 %RH
0–100 %RH, non‑condensing
Type 4X: Dust-tight. Protected from corrosion
and hose-directed water.
IP66: Dust-tight. Protected from powerful water
jets from any direction.
IP67: Dust-tight. Protected from the effects of
temporary immersion in water under
standardized conditions of pressure and time.
1)
2)
/ NPT ½"
M212931EN-C