5.7. Memory full
Pay attention that the maximum memory capacity is not
exceeded. When the memory is full, the old values are
automatically Overwritten with new ones. When mem-
ory is full , the display shown 1 second as follows to
remind you " memory full "
5.8. Memory– cancellation of all measurements
Attention!
Before you delete all readings stored in the memory,
make sure you will not need refer to the readings at
a later date. Keeping a written record is prudent and
may provide additional information for your doctor's
visit. In order to delete all stored readings, depress the
MEMORY button for at least 5 seconds, the display
will show the symbol «CL» and then release the button
. to permanently clear the memory, Press the MEMO-
RY button while «CL» is flashing. 3 short beep sounds
will be heard to indicate deletion of stored readings.
6.Error messages /malfunctions
If an error occurs during a measurement, the measurement is discontinued and
a corresponding error code is displayed.
Error No.
Possible cause(s)
ERR 1
No pulse has been detected.
ERR 2
Unnatural pressure impulses influence the measurement result.
Reason: The arm was moved during the Measurement (Artefact).
ERR 3
The inflation of the cuff takes too long. The cuff is not correctly
seated.
ERR 5
The measured readings indicated an unacceptable difference
between systolic and diastolic pressures. Take another reading
following directions carefully. Contact you doctor if you contin-
ue to get unusual readings.